Basic Maintenance Page 21
Fig. 19: Pawl drive assembly, pulled back from front plate.
8. Using the pins on the front, unscrew the load lock door from its post
(counterclockwise).
9. Clean the load lock O-ring and the flat side of the door with a clean lab
wipe and alcohol (Fig. 20).
Fig. 20: Load lock seal, and load lock seal door.
Load lock seal Load lock door, seal side