BIO RAD 200/2.0 Instruction Manual

Browse online or download Instruction Manual for Power supply units BIO RAD 200/2.0. BIO RAD 200/2.0 Instruction manual User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 12
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
Model 200/2.0
Power Supply
Instruction
Manual
Catalog Numbers
165-4761 and 165-4762
For Technical Service Call Your Local Bio-Rad Office or in the U.S. Call 1-800-4BIORAD (1-800-424-6723)
Page view 0
1 2 3 4 5 6 ... 11 12

Summary of Contents

Page 1 - Instruction

Model 200/2.0Power SupplyInstructionManualCatalog Numbers165-4761 and 165-4762For Technical Service Call Your Local Bio-Rad Office or in the U.S. Call

Page 2 - Table of Contents

8. Sponge pads can not dissipate heat as effectively as Scoth Brite®gel pads. For high power,high intensity applications, it is advisable to use gel p

Page 3 - Warranty

Section 4Troubleshooting GuideCondition Cause Remedy1. With the power supply 1a. Power cord is not properly 1a. Check connections.on, the LED fails to

Page 4 - Section 1

Eastern Regional Office, 85A Marcus Dr., Melville, New York 11747 • Phone (516) 756-2575 • Fax (516) 756-2594Australia, Bio-Rad Laboratories Pty Limit

Page 5 - 1.3 Specifications

Table of ContentsPageSection 1 General Information...11.1 General In

Page 6

NoteTo insure best performance from the Model 200/2.0 Power Supply, become fully acquainted withthese operating instructions before use.WarrantyBio-Ra

Page 7 - Operation

1Section 1Introduction1.1 General InformationBio-Rad’s Model 200/2.0 microprocessor controlled power supply provides constantvoltage to dual output j

Page 8 - 2.2 Time Mode Operation

1.3 SpecificationsOutput voltage 5-200 volts DC, continuously adjustableOutput current 0.01-2.0 A, continuously adjustableOutput terminals Floating d

Page 9 - Blotting Applications

1.4 Diagram of Front Panel Membrane Control Pad1.5 Description of Membrane Control Pad1. LED display Displays the current limit, voltage, and timer

Page 10

Section 2Operation2.1 Standard Mode OperationGeneral Application: The standard mode is used in blotting applications and in nucleicacid electrophore

Page 11 - Troubleshooting Guide

Turn the Power Supply OffThere are three different ways to turn the power supply off.1. Press VOLTS. The red indicator light should be on. Press LOWER

Page 12 - Life Science

63. Press LIMIT to view the set current limit. The red indicator light above the button shouldbe on, and the set limit will be shown on the LED displa