INSTRUCTION MANUALCatalog Numbers165-1745165-1746MODEL 583 GEL DRYERMODEL 583GEL DRYER
7Gel Preparation Techniques3.0 GEL PREPARATION TECHNIQUESTechniques for drying the following gels are discussed in this section:• Sequencing Gels• Acr
8Gel Preparation TechniquesTable 4Drying GelsStep Procedure1. For gels with radioactive label,place a dry sheet of filter paper ontop of the porous Ge
9Gel Preparation TechniquesTable 4 (continued)Drying GelsStep Procedure6. Program the drying cycle.DescriptionFor Sequencing gels:a. Select the Sequen
10Gel Preparation TechniquesTable 4 (continued)Drying GelsStep Procedure7. Press Start to begin the dryingcycle.8. When the cycle is complete, check t
11Suggested Drying Times4.0 SUGGESTED DRYING TIMESGel drying time is a function of temperature, vacuum, gel porosity, and gel thickness. With the exce
12Maintenance and Troubleshooting5.0 MAINTENANCE AND TROUBLESHOOTINGGel Dryer maintenance consists of:• Cleaning and replacing the porous Gel Support•
13Maintenance and TroubleshootingTable 6Troubleshooting GuideIndication ProblemNo power; front panel No AC power to the gel display is blank. dryer.G
14Maintenance and TroubleshootingTable 6 (continued)Troubleshooting GuideIndication ProblemVacuum seal is lost Vacuum leak in gel dryer.during opera
15Maintenance and TroubleshootingTable 6 (continued)Troubleshooting GuideIndication ProblemWet spot appears on The vacuum pump wasthe gel at the loc
16Vacuum Source OptionsAPPENDIX A. VACUUM SOURCE OPTIONSA.1 VACUUM SOURCE REQUIREMENTSVacuum is commonly measured in inches of mercury (“ Hg) or milli
TABLE OF CONTENTSSafety ...
17SpecificationsAPPENDIX B. SPECIFICATIONSDryer body PolycarbonateDrying surface 35 x 45 cmWeight 9.8 kgDimensions 11 x 55 x 50 cmGel support plate Po
18Warranty and Ordering InformationAPPENDIX C. WARRANTY AND ORDERING INFORMATIONThe Model 583 Gel Dryer is warranted for 1 year against defects in mat
19Warranty and Ordering InformationORDERING INFORMATIONCatalogNumber DescriptionGel Drying Systems165-1789 HydroTech Gel Drying System, 100/120 V, inc
M1651740 Rev HLife ScienceGroupWeb site www.bio-rad.com Bio-Rad Laboratories Main Office 2000 Alfred Nobel Drive, Hercules, CA 94547, Ph. (510) 741-1
iSafetySAFETYCaution/WarningDisconnect the Gel Dryer before servicing. No user-serviceable parts areinside. Refer servicing to Bio-Rad service personn
1Introduction1.0 INTRODUCTIONThe Model 583 Gel Dryer is designed to dry sequencing and protein gels fast! With three pre-programmeddrying cycles, the
2Introduction1.2 PHYSICAL DESCRIPTIONTable 1.Physical FeaturesFeature Description1. Dryer Control Panel The front panel controls are used to program d
3IntroductionTable 1 (continued)Physical FeaturesFeature DescriptionTEMP: Drying temperatures can be set between 50° C and 90° C. To set adrying tempe
4Set Up and Operation2.0 SET UP AND BASIC OPERATION2.1 SET UPTable 2 discusses the electrical and vacuum connections required for the Model 583 Gel Dr
5Set Up and Operation2.2 OPERATIONOnce the drying parameters are set, they are retained in memory until changed by the user or until power isturned of
6Set Up and OperationTable 3 (continued)Gel Dryer OperationStep Procedure7. Start the drying cycle.8. Adjust the temperature or time at any time durin