EXQuest
Spot Cutter
Instruction Manual
Catalog numbers
165-7200 EXQuest Spot Cutter
165-7201 EXQuest Spot Cutter With Computer
On the Web at http://www.bio-rad.com
For Technical Service, Call Your Local Bio-Rad Office or, in the US, Call 1-800-4BIORAD (1-800-424-6723)
Page view 0
1 2 3 4 5 6 ... 80 81

Summary of Contents

Page 1 - Instruction Manual

EXQuest™Spot Cutter Instruction ManualCatalog numbers165-7200 EXQuest Spot Cutter 165-7201 EXQuest Spot Cutter With Computer On the Web at http://ww

Page 2

•Gel Tip Wash Station. The gel tip wash station is located on the left end of the wash station block. Itconsists of a gel tip wash port, a waste basin

Page 3

Figure 1-8. Supported microplate formats.The microplate rack accommodates plates from a variety of vendors that include Corning, Whatman,Greiner, AxyG

Page 4

1.2.9 Sensors The spot cutter includes a series of sensors to protect the operator and ensure the proper operation of theinstrument.•Front Door Sensor

Page 5 - 1.0 INTRODUCTION

1.2.11 Software The EXQuest spot cutter is controlled through Bio-Rad Discovery Series™software PDQuest (7.4.0 orgreater) and Quantity One (4.6 or gre

Page 6

2.0 SYSTEM INSTALLATIONBefore the EXQuest™spot cutter can be used, the software must be installed and the instrument set up andcalibrated. At the time

Page 7

CautionThe EXQuest spot cutter must be lifted, by two or more individuals, from its shipping pallet using theremovable lifting handles attached to the

Page 8 - Gel Cutting Head

SYSTEM INSTALLATION2-32.3 SYSTEM LEVELING To level the EXQuest spot cutter:1. Place the spot cutter on a lab bench that is sturdy and level.2. Loosen

Page 9

SYSTEM INSTALLATION2-42.5 MICROPLATE RACK INSTALLATIONThe microplate rack is installed to the left of the gel tray as follows:1. Locate the two micro

Page 10 - INTRODUCTION

Figure 2-5. Installed microplate rack.2.6 CABLE SETUPBefore the spot cutter can be started, the power cord and USB cable must be connected.1. Connect

Page 11 - Figure 1-9. Robotics

SYSTEM INSTALLATION2-6Figure 2-6. Spot cutter USB ports and power entry module.2.7 CUTTING HEAD INSTALLATIONThere are two types of cutting heads avai

Page 12

TABLE OF CONTENTSTABLE OF CONTENTSSAFETYSECTION 1 INTRODUCTION ...

Page 13 - Manual Excision Tool

2-7SYSTEM INSTALLATIONTo install or change the gel cutting head1. Open the front door by lifting it all the way up and then lowering it until it latch

Page 14 - 2.0 SYSTEM INSTALLATION

SYSTEM INSTALLATION2-85. Insert three clean 1.5 ml capless tubes into the membrane tip wash port slots, located on the washstation’s right side. Fill

Page 15 - SYSTEM INSTALLATION

SYSTEM INSTALLATION2-97. Read and accept the License Agreement.8. Click Next on the Read Me Screen.9. Enter your Full Name and Organization and then c

Page 16 - Figure 2-1. Spot cutter leg

3.0 EXQUEST CALIBRATION WIZARDBefore you can use the spot cutter for the first time, you must calibrate it. The calibration wizard helps you toquickly

Page 17 - Widely spaced tabs

4. Click Setup spot cutter… to start the "EXQuest Calibration Wizard". The screen that will appeardepends on the current calibration state o

Page 18

EXQUEST CALIBRATION WIZARD3-3Figure 3-4. “EXQuest Calibration Wizard” screen – calibrated.From this screen you may• Calibrate the camera with the curr

Page 19

EXQUEST CALIBRATION WIZARD3-4Figure 3-5. Camera focusing dialog.6.Head Height Calibration. The Head Height Calibration dialog is used to set the cutti

Page 20

EXQUEST CALIBRATION WIZARD3-5Figure 3-6. “Head Height Calibratioin” screen.7.Flat Field Calibration. The flat field calibration step corrects for inho

Page 21

EXQUEST CALIBRATION WIZARD3-68.Lens Calibration. The lens calibration step corrects for image distortion due to the camera lens. Toperform the lens ca

Page 22

Figure 3-9. System position calibration.Manual Calibration a. If calibrating the membrane cutting head, install the plastic protective sleeve over th

Page 23 - EXQUEST CALIBRATION WIZARD

SECTION 7 LANE VOLUME EXCISION TOOL...7-17.1 Setting Up a Cut Run

Page 24

Figure 3-10. Plate position calibration: plate selection.b. Use the Left/Right and Front/Back slider to center the cutting tip over well A1, as shown

Page 25 - Camera Focus

d. Use the Raise/Lower slider to adjust the tip height. Typically, the gel tip should be placed about0 to 1 mm into the well and the Membrane cutting

Page 26 - Head Height Calibration

Figure 3-13. Wash station position calibration.b. When the tip is centered, check Lower head to adjust wash height to lower the cutting tip intothe we

Page 27 - Flat Field Calibration

Figure 3-14. Calibration complete! screen.EXQUEST CALIBRATION WIZARD3-11

Page 28 - System Position Calibration

4.0 GEL AND BLOT PREPARATIONGel and blot preparation is an important part of setting up a successful cut run. Failure to prepare gels andblots properl

Page 29 - Plate Position Calibration

4.3 BLOT PREPARATION1. Store PVDF and nitrocellulose blots in a container that keeps them protected and hydrated untilneeded for spot cutting.2. Remov

Page 30

5.0 MANUAL EXCISION TOOLThe Manual Excision Tool is a general-purpose excision tool included in both the PDQuest™and QuantityOne®software. It has the

Page 31

Figure 5-2. Manual excision tool.7. Close the spot cutter front door and click Acquire Image to take an image of the gel tray contents.Do not open the

Page 32 - Calibration Complete

12. Optional. Check Hydrate in the "Hydration Options" panel to automatically hydrate the gel during thecut run. Specify how often it is to

Page 33

17. Click Begin/Resume to open the "Plates for Cut Run" dialog (see Section 8.9) and then enter theplate name and barcode/plate ID in their

Page 34 - 4.0 GEL AND BLOT PREPARATION

SAFETY INFORMATIONRead the manual before operating the EXQuest Spot Cutter. For technical assistance, contact your local Bio-Rad representative or, in

Page 35 - GEL AND BLOT PREPARATION

Figure 5-5. Import excision coordinate list.3. If the first line of your file is header information, check Skip first line of file.4. Enter the column

Page 36 - 5.0 MANUAL EXCISION TOOL

MANUAL EXCISION TOOL5-6Figure 5-6. Confirm cuts pop-up.2. Read the instructions on the "Image Microtiter Plate" screen.Figure 5-7. “Image Mi

Page 37

5.3.2 Identifying Re-cuts from a Gel Image1. When the cut run is finished, return to the "Manual Excision Tool" screen and click Confirm Cu

Page 38 - MANUAL EXCISION TOOL

6.0 ANALYSIS SET EXCISION TOOLThe Analysis Set Excision Tool is designed to use the advanced capabilities of the PDQuest™software increating cut reque

Page 39 - Gel Imaging

ANALYSIS SET EXCISION TOOL6-24. Click Next to continue. The wizard screen that appears will depend on whether a cut request list hasalready been creat

Page 40

From the "Analysis Set with Unfinished Cuts" screen (Figure 6-4):– Select Resume cut run and finish these cuts and then click Next, to finis

Page 41

Figure 6-6. “Select Gels to Cut” ScreenFrom the "Select Gels to Cut" screen, select one or more gels to cut. Multiple gels may beselected by

Page 42

a. From the "Automatically Select Gels to Cut" screen, set the Tolerance. The tolerance value isused to determine the spot quantity range th

Page 43 - ANALYSIS SET EXCISION TOOL

•PVDF: This option is available only when the spot cutter is in membrane cutting mode, and isused to specify that either a PVDF or Nitrocellulose memb

Page 44

Figure 6-9. Analysis Set Excision Control Tool•Analysis set name: This text field displays the name of the selected analysis set.•Analysis set spot co

Page 45 - Manual Gel Selection

1.0 INTRODUCTION1.1 OVERVIEWThe EXQuest™spot cutter is a precision instrument that accurately locates and excises protein spots andbands from one and

Page 46 - Automatic Gel Selection

Figure 6-10. Place Gel/Blot on Spot Cutter dialog box: Single Gel PlacementFigure 6-11. Place Gel/Blot on Spot Cutter dialog box: Multiple Gel Placeme

Page 47

14. Align the spots on the cutter image (right side of Figure 6-12) with spots in the analysis set image(left side of Figure 6-12).Figure 6-12. Align

Page 48

Figure 6-14. Align Analysis and Cutter Images: Step 2 – Perform Alignment Panel•Automatic Alignment. Automatic alignment uses a Gaussian fitting algor

Page 49

Figure 6-16. Align Analysis and Cutter Images: Jump and ConfirmThe following image options are available in this window:•Display cropped image. Check

Page 50

Figure 6-17. Align Analysis and Cutter Images: Cut Overlaps15. Click Proceed to hydrate the gel or blot (if hydration is turned on) and open the "

Page 51

17. When the run is finished, you will be prompted to place the next gel on the spot cutter, if you arecutting multiple gels. Place the gel on the gel

Page 52

ANALYSIS SET EXCISION TOOL6-143. Check Analysis Set Excision Tool advanced user mode in the "Spot cutter" panel and then clickOK.4. Make sur

Page 53

Figure 6-23. Step 2 – Select Gels: Select from ListAuto-selecta. Choose Auto-select to let PDQuest automatically select the best gel(s) to cut based o

Page 54

12. Optional. Click the Review tab and review the cut selections. The Review tab displays the gel spotsby Standard Spot Number (SSP) along with their

Page 55

•Plastic-backed Gels. Select this option to use a protocol optimized for cutting plastic-backedgels. When this option is used, gel plugs are cut and t

Page 56

1.2 SYSTEM DESCRIPTIONThe EXQuest spot cutter is a fully integrated system controlled through PDQuest™and Quantity One®software. It consists of an enc

Page 57

7.0 LANE/VOLUME EXCISION TOOLThe Lane/Volume Excision Tool is designed to provide a simple and efficient method for cutting lanes andbands from 1D gel

Page 58

Tips for cutting membranes– Fill microplate wells with DI water to within 1 to 2 mm of the top. In order to release the cutmembrane into a microplate

Page 59 - Glass-backed gels

Figure 7-3. Lane/Volume Excision Tool image with lanes3. Click Stretch Frame and then click and drag the frame anchors to fit the frame over the gel.4

Page 60 - 7.0 LANE/VOLUME EXCISION TOOL

6. Adjust Horizontal slider to control the horizontal cut spacing. Use the zoom tool to see the effect ofchanging the Horizontal control.Figure 7-5. T

Page 61 - LANE/VOLUME EXCISION TOOL

Figure 7-7. Lane/Volume Excision Tool: Display Excisions4. Review the "Run Info" panel (see Section 8.8) for the number of cuts selected, th

Page 62

EXCISION TOOL CONTROLS8-18.0 EXCISION TOOL CONTROLSThis section describes the excision tool controls that are common between the Manual Excision Tool,

Page 63 - Right, Horizontal = 0

EXCISION TOOL CONTROLS8-28.3 TRANSFORM DIALOG BOXThe "Transform" dialog box is used to adjust the image brightness and contrast. It contain

Page 64

EXCISION TOOL CONTROLS8-3Image Panel•Max: This text box displays the intensity of the image's most intense pixel. Image units aredetermined by th

Page 65 - 8.0 EXCISION TOOL CONTROLS

EXCISION TOOL CONTROLS8-4•Material: Use this drop-down menu to select the gel type that will be cut, e.g., "Free gel", "Plastic-backed

Page 66 - Preview Window

EXCISION TOOL CONTROLS8-5•Define hydration area. To define the cutting deck area to hydrate, click this button and then click anddrag a box around the

Page 67

INTRODUCTION1-3Figure 1-3. Light enclosure.1.2.2 Imaging SystemThe gel and blot imaging system consists of a CCD camera (1600 x 1200 pixels, 12 bit AD

Page 68

EXCISION TOOL CONTROLS8-68.9 PLATES FOR CUT RUNFigure 8-9. “Plates for Cut Run” dialog boxThe "Plates for Cut Run" dialog is used to identi

Page 69

EXCISION TOOL CONTROLS8-78.10 LOAD PLATE ON CUTTER DIALOGFigure 8-10. “Load Plate on Cutter” dialog boxThis dialog helps the operator correctly place

Page 70

ACCESSORIES9-19.0 ACCESSORIES, SUPPLIES, REPLACEMENT PARTSCatalogNumber Description 165-7208 EXQuest Cutting Sheets (S), pkg of 15165-7205 EXQuest Ca

Page 71

10.0 MAINTENANCE10.1 ROUTINE MAINTENANCE• The gel tray should be cleaned after use with a damp nonabrasive cloth. Mild detergents may beused. Do not u

Page 72 - ACCESSORIES

MAINTENANCE10-2Figure 10-2. UV EPI Lamp Partially Removed3. Place the lamp assembly upside down on a flat surface and then rotate each fluorescent tub

Page 73 - 10.0 MAINTENANCE

MAINTENANCE10-3Figure 10-4. Spot cutter fuse

Page 74 - MAINTENANCE

TROUBLESHOOTING11-111.0 TROUBLESHOOTINGSymptom Cause SolutionComputer does not USB cable not Ensure that both the USB cable and power cord communi

Page 75 - Figure 10-4. Spot cutter fuse

TROUBLESHOOTING11-2Symptom Cause SolutionHydration and clamping Excess water on the gel tray causes backed gels toslide when the gel plug is being scr

Page 76 - 11.0 TROUBLESHOOTING

TROUBLESHOOTING11-3If all the settings seem to be correct, contact Bio-Rad for further troubleshooting. For technical assistancewith the system includ

Page 77 - TROUBLESHOOTING

12.0 WARRANTYEXQUEST™SPOT CUTTER WARRANTY INFORMATIONThis warranty statement may vary outside of the continental United States. Please contact your l

Page 78

INTRODUCTION1-4Figure 1-4. Gel tray with clamps.1.2.4 Cutting HeadsThe EXQuest spot cutter supports both gel and membrane (blot) cutting heads. The ge

Page 79 - 12.0 WARRANTY

SYSTEM RECORDSSerial NumberDate of PurchasePurchase OrderNumberNotesService RecordDate Description of Problem Name SignatureWARRANTY12-2

Page 80

Life ScienceGroup00-000 0000 Sig 1204Bulletin 0000 US/EG Rev ABio-Rad Laboratories, Inc.Web site www.bio-rad.com USA (800) 4BIORAD Austr

Page 81 - Laboratories, Inc

INTRODUCTION1-5•Membrane Cutting Head. The membrane cutting head is an optional accessory that is used forcutting PVDF and nitrocellulose membranes. A