Page 1 - Instruction Manual

PowerPac™ Universal Power SupplyInstruction ManualCatalog Number164-5070For Technical Service Call Your Local Bio-Rad Office or in the U.S. Call 1-800

Page 2 - List of Figures

Step Procedure Description8. Set the key chirp mode Press the CHIRP soft key to turn key chirp onor off.KEY CHIRP = OFF: No sound is produced when a

Page 3 - Caution/Warning

Step Procedure Description4. Enter limits for the Use the soft keys to select a non-constant non-constant parameter and the keypad to enter its valu

Page 4 - Introduction

Step Procedure Description2. Select the programming On the Home screen, press the METHODSmode soft key to display the Method List screen.3. Create a

Page 5 - 1.2 Unpacking

Step Procedure Description7. Enter non-constant Use the soft keys to select a non-constant parameter limits for parameter and the keypad to enter it

Page 6

3.3.2 Edit (or View) and Run a Saved MethodThis section describes how to edit, view and run a saved method.Step Procedure Description 1. Set up the p

Page 7 - Control features

Step Procedure Description 6. Set the non-constant Use the soft keys to select a non-constant parameter limits for parameter and the keypad to enter

Page 8 - Setup and Operation

Step Procedure Description 11. View the run The "Run Completed" or "Run Terminated" information screen is displayed at the end of

Page 9

Step Procedure Description 5. Set the non-constant Use the soft keys to select a non-constant parameter limits for parameter and the keypad to enter

Page 10 - 3.2 Manual Mode Operation

Section 4Operation with a Personal Computer and PDA The PowerPac Universal is capable of transmitting methods to and from aPDA, and transmitting run d

Page 11 - 3.3 Method Mode Operation

Step Procedure Description2. Open a method There are three ways to open a method: a. Press New Method to create and open anew method.b. Select a met

Page 12 - MANUAL METHODS

Table of ContentsPageSafetySection 1 Introduction ...11.1 Overview...

Page 13

Step Procedure Description4. Edit a step a. Set the Constant mode: voltage (V), cur-rent (A) or power (W).b. Enter the constant and non-constant

Page 14

Step Procedure Description2. Receive data from the Press Get Data From PowerPac to request PowerPac Universal and receive the data file from the Pow

Page 15

Section 5Maintenance and Troubleshooting5.1 MaintenanceThe PowerPac Universal requires little maintenance to assure reliable operation. To clean the

Page 16

If a power failure occurs during a run the following screens will be displayed. 1. Displayed if PFd = ONMORE is displayed on the Run Completedscreen w

Page 17

5.2.4 Error MessagesError Message Cause SolutionERROR STOP CODE 01 1. Electrophoresis cell 1. Make sure all electrical NOLOAD DETECTED not connec

Page 18 - PowerPac Remote Control

Error Message Cause SolutionERROR STOP: CODE 09 1. Loose output 1. Verify all electrical RAPID CHANGE IN R connections leading connections.CORREC

Page 19

5.3 Replacing a FuseIf the power supply is plugged into a working outlet, with the power switch inthe ON position and there is no display, lights or

Page 20

Appendix ASpecificationsOutput specifications: 500 V, 2500 mA, 500 WOutput range (programmable): 10–500 V, fully adjustable in 1 V increments10–2500 m

Page 21 - Get Data From PowerPac

Appendix BWarranty and Ordering InformationWarrantyThe PowerPac Universal is warranted for 1 year against defects in materialsand workmanship. If any

Page 22 - Section 5

Life ScienceGroup 0104 Sig 1103Bulletin 3085 US/EG Rev ABio-Rad Laboratories, Inc.Web site www.bio-rad.com USA (800) 4BIORAD Australia 0

Page 23

SafetyCaution/WarningPowerPac power supplies use high output voltages that are electrically isolated from earthground to minimize the risk of electric

Page 24 - 5.2.4 Error Messages

Section 1Introduction1.1 OverviewThe PowerPac Universal power supply is designed to provide constant voltage,current or power for a wide range of ele

Page 25

2The PowerPac Universal has the following features:• Constant voltage, constant current, or constant power operation with automaticcrossover• LCD scre

Page 26 - 5.3 Replacing a Fuse

After unpacking the PowerPac Universal, remove the plastic film from thetranslucent green top case. The plastic film may leave a residue. If so, clean

Page 27 - Specifications

Section 2Control featuresFig. 4. Front Panel.Key DescriptionRun/Pause Starts or pauses a run. When paused, the run parameters forthe current or subseq

Page 28 - Appendix B

Section 3Setup and Operation3.1 SET UPThis section describes how the PowerPac Universal power supply is set up andconnects to an electrophoresis cell

Page 29 - Laboratories, Inc

Step Procedure Description4. Set the Power Failure Use the PFd soft key to set the Power Failure detection mode detection mode:PFd = OFF: Turns powe