Step Procedure Description
11. View the run The "Run Completed" or "Run Terminated"
information screen is displayed at the end of a run. The
date and time, the final running parameter
values, the total elapsed time in hours and
volt-hours are displayed on the screen. Press
the RERUN soft key to repeat the run or the
EXIT soft key to display the Method Step List.
After the run is finished, the run data may be
transmitted to a PDA or PC (see Section 4.2.2).
3.3.3 Edit a Paused Run
This section describes how to edit a paused run.
Step Procedure Description
1. Pause the run Press the PAUSE soft key or to pause a
run that is in progress.
2. Edit the method The pause window displays the current run
parameters as well as the date and time that
the run was paused.
Press the EDIT soft key or the EDIT key to
modify the method. The CONTINUE soft key
or button can be used to resume the run.
3. Select a method step Press the up and down arrow keys to select
the step to edit or add. An asterisk (*) indicates
the selected step.
Press the EDIT key to modify the selected
step.
Press the CONTINUE soft key or to
resume the run when finished editing the
method.
4. Set the constant Use the soft keys to select the constant
parameter for the parameter and the keypad to enter its value.
current step Press the EDIT key when finished. The
key can be pressed at any time to resume the
run. Note that the non-constant parameter
values are listed on the right side of the screen
and the time mode at the bottom.
ConstV: Used to run with constant voltage.
ConstA: Used to run with constant current.
ConstW: Used to run with constant power.
13