2 | The NGC Instrument
66 | NGC Chromatography Systems and ChromLab Software
Single-Wavelength UV Detector
Detailed Attributes
Two integrated LED UV bulbs
LED operating usage: approximately 5,000 hours
Note: If the reference output is below 0.9 volts you might need to replace
the LED bulbs depending on the requirements of the experiment.
For information about changing the LED bulbs, see Replacing the Single-
Wavelength UV Detector LED on page 201.
L
EGEND
1 Conductivity monitor 2 LED display
3 UV flow cell 4 UV detector housing
5 Point-to-Plumb™ indicator