2 | The NGC Instrument
68 | NGC Chromatography Systems and ChromLab Software
Multi-Wavelength UV/Vis Detector
Detailed Attributes
Contains one tungsten and one deuterium lamp
Lamp operating usage: approximately 2,000 hours for the tungsten bulb as
well as the deuterium bulb
For information about changing the UV lamps, see Replacing the Multi-
Wavelength UV/Vis Detector Lamps on page 209.
Tip: To conserve lamp life, in Manual mode open the Multi Wave Detector
dialog box and turn off the UV/Vis lamps when the instrument is not in use.
L
EGEND
1 Conductivity monitor 2 UV/Vis detector flow cell
3 UV/Vis detector housing