Detectors
Instrument Guide | 6 7
Monitors UV absorbance one wavelength at a time
255 nm
280 nm (default)
Tip: Use ChromLab to change the absorbance wavelength values. For
more information, see the NGC Chromatography Systems and ChromLab
Software User Guide.
UV absorbance range: 0–3,000 mAU
UV linearity: ±5%
Three interchangeable flow cells with path lengths of 2 mm, 5 mm (default),
and 10 mm; see Table 4 on page 69 for specific information
Note: Use the preparative 2 mm flow cell to measure high concentrations
of protein. The path lengths of the analytical 5 mm and 10 mm flow cells
are longer and are appropriate for measuring lower concentrations of
protein.
For information about changing the flow cells, see Replacing the UV Flow Cell
on page 196.