Connecting Peripheral Equipment
Installation Guide | 5 7
3. Insert system cable 17 that ships with the fraction collector into an
INSTRUMENT BUS connector port.
4. Insert the other end of the cable into the INSTR. BUS connector on the back of
the communication adaptor.
Tip: The RX and DX diodes are communication indicators. You see the
lights flash as data transfer between the fraction collector and the NGC
instrument. The COMM port is not used to connect to the fraction
collector
.
5. Locate the peripheral cable that ships with the communication adaptor.